Home>Product>Lock in Amplifier
TW4803 DSP Lock-In Amplifier

- 1 mHz to 102 kHz frequency range

- 1 nV to 1 V full-scale sensitivity

- Time constants from 10 µs to 3 ks

- > 120 dB dynamic reserve

- Automatic adjustment

- Multiple-harmonic measurement

- FFT spectral analysis

- 5.6 inch color TFT-LCD screen, single or dual display

- Support remote control, embedded a free LabVIEW program


  • Features
  • Specifications
  • Ordering Information
  • Download

Maxwellon TW4803 Digital Lock-in Amplifier provides an excellent performance within its bandwidth from 1 mHz to 102 kHz. With the advantage of the latest digital signal processing technology and high-precision 24-bit ADC, TW4803 can easily detect the phase and the magnitude of weak signals overwhelmed by various large noises. The performance of TW4803 is as good as other lock-in amplifiers brands, even better in some certain parameters, such as measurement accuracy, SNR, dynamic reserve. Otherwise, TW4803 integrates some special functions like multiple harmonic measurement and FFT, which meets the needs of scientific research and industrial application well.

Signal Channel

Voltage Input Mode

Single-ended or Differential

Full-scale Sensitivity

1 nV to 1 V in a 1-2-5 sequence

1 fA to 1 µA

Current Input

106 or 108 V/A

Impedance

• Voltage: 10 MΩ // 25 pF, AC or DC coupled

Current: 1 kΩ to virtual ground

C.M.R.R

> 100 dB to 10 kHz, decreasing by 6   dB/oct

Dynamic Reserve

> 120 dB

Gain Accuracy

0.2% typ, 1% max

Voltage Noise

5 nV/Hz at 997 Hz

Current Noise

15 fA/Hz at 97 Hz

13 fA/Hz at 997 Hz

Line Filters

50/60 Hz and 100/120 Hz

Gounding

BNC shield can be grounded or floated via 10   kΩ to ground

Reference Channel

Input

• Frequency range:  1 mHz to 102 kHz

• Reference input: TTL or Sine

Input impedance: 1 MΩ//25 pF

Phase

• Resolution: 0.01°

Absolute phase error: < 1°

Relative phase error: < 0.01°

Orthogonality: 90° ± 0.001°

Phase noise: (Internal   ref.) Synthesized, <0.0001°rms at 1 kHz;  (External ref.) 0.001°rms at 1 kHz (100 ms time constant, 12 dB/oct)

Drift: <0.01°/℃ below 10 kHz, <0.1°/ above 10 kHz

Harmonic Detection

2F, 3F, nF to 102 kHz  (n<32767)

Acquisition Time

• Internal ref.: instantaneous acquisition

• External ref.: (2 cycles + 5 ms) or 40 ms, whichever is larger

Demodulator

Stability

Digital output: no   zero drift on all setting

• Display: no zero drift   on all setting

• Analog output: <5 ppm/ for all dynamic reserve settings

Harmonic Rejection

-90 dB

Time Constant

10 µs to 3 ks   (<200 Hz)

10 µs to 30 s   (>200 Hz)

(6, 12, 18, 24 dB/oct rolloff)

Synchronous Filters

Available below 200 Hz (18, 24 dB/oct rolloff)

Internal  Oscillator

Frequency

Range: 1 mHz to 102   kHz

Accuracy: 2 ppm + 10 µHz

Resolution: 1 mHz

Distortion

-80 dBc (f < 10   kHz)

• -70 dBc (f > 10 kHz)

Amplitude

0.001 to 5 Vrms

Accuracy

1%

Stability

50 ppm/

Output

Sine output on front panel, and TTL sync   output on rear panel

General Specifications

Interface

RS-232 to USB interface, and IEEE-488 interface (optional).

CH1 and CH2 Outputs

• Function: output X, Y,   R, θ

Output voltage: ±10 V full scale, 30 mA max output current

Update rate: 312.5 kHz

Screen

5.6 inch, 640×480 TFT

Screen Format

Single or dual display

Display Quantities

Each display shows one trace, traces can be   defined as X,Y,R,θ

Display Types

Numerical form, bar graph, polar plot and   strip chart

Power Requirement

Voltage: 220 240 VAC, 100 120 VAC   (optional)

• Frequency: 50/60 Hz

Power: 30 W

Dimension

448 (W) × 148 (H) × 513 (D) mm  (with feet)

Weight

11 kg


Ordering Information

Part No.

Name

Description

TW4803

Digital Lock-in Amplifier

1 mHz to 102 kHz, 1 nV to 1 V full-scale sensitivity


Stay Informed